医用CT剂量指数测量结果的不确定度评定

2251    2016-01-23

免费

全文售价

作者:高铁, 李国琛

作者单位:温州市计量技术研究院, 浙江温州 325001


关键词:医用诊断计算机断层摄影装置(CT)X射线辐射源; 剂量指数; 测量不确定度


摘要:

为了使医用诊断计算机断层摄影装置(CT)X射线辐射源在模体中的剂量指数测量结果的不确定度评定合理,基于用带有100mm长杆电离室的积分剂量仪测量CT剂量指数的数学模型,对其测量不确定度来源进行分析,列出了方差、灵敏系数及标准不确定度的计算公式,从而给出了CT剂量指数测量的合成标准不确定度和扩展不确定度。


Uncertainty analyses of computed tomography dose index

GAO Tie, LI Guo-chen

Wenzhou Institute of Measurement Technology, Wenzhou 325001, China

Abstract: In order to make the uncertainty of the dose index measurement in the head phantom reasonable for the medical diagnostic X-ray radiation sources of the computed tomography (CT), the measurement uncertainty components were analyzed according to the mathematical model for the computed tomography dose index test method with a 100 mm ionization chamber. And the calculated formula for square-error, sensitivity coefficient and standard deviation were listed in the article. As the results, the standard measurement uncertainty and its expanded uncertainty of CT dose index were presented.

Keywords: Medical diagnostic X-ray radiation source for computed tomography; Dose index; Uncertainty

2008, 34(6): 21-23,43  收稿日期: 2008-3-19;收到修改稿日期: 2008-6-4