DXA骨密度标准体模的研究

2680    2016-01-16

免费

全文售价

作者:鄢铃, 张园月, 魏川, 黄扬, 林滔

作者单位:中国测试技术研究院, 四川成都 610021


关键词:DXA仪; 骨密度; 标准体模


摘要:

文中对双能X线骨密度仪(DXA)的腰椎骨密度标准体模进行了研究。该体模由水等效塑料和骨等效羟磷酸钙组成,其外尺寸为186(侧位)150(前后位)156(轴向),模拟了DXA仪在测量腰椎骨时的前后位(AP)测量状态。体模的水等效材料的CT值在15 Hu内;羟磷酸钙(HA)骨样品有三种密度,分别是0.5 g/cm2,1.0 g/cm2和1.5 g/cm2,对骨密度仪进行检定和线性校正。体模的性能试验使用了Ge-Lunar厂家生产的双能X射线骨密度仪,测得重复性小于2%,测量结果误差在2%~4%之间,与同类模体欧洲体模(ESP)作了比较,性能基本一致。该标准体模的研究,对骨密度量值的准确和统一,具有重大的意义。


Research on standard phantom of DXA bone density

YAN Ling, WEI Chuan, HUANG Yang, ZHANG Yuan-yue, LIN Tao

National Institute of Measurement and Testing Technology, Chengdu 610021, China

Abstract: The spine bone density standard phantom was researched with the dual-energy x-ray absorptiometry (DXA).The phantom was made of water equivalent polyethylene-based plastic and bone equivalent calcium hydroxyapatite, and its physical size was 186(lateral)×150(AP)×156(axis).The AP status of measured spine bone was simulated by DXA.Inside the phantom, the CT values of the water equivalent plastic were within±15Hu(Ex:10 Kev~100 Kev).There were three densities of calcium hydroxyapatite(HA) in bone samples: 0.5 g/cm2, 1.0 g/cm2 and 1.5 g/cm2 respectively.The bone density was calibrated with the regression line.The property experiment of the phantom was performed by a DXA machine made by Ge-Lunar.The results showed the relative standard deviations of measurements were less than 2%, and measuremental errors were from 2% to 4%.The comparison results show that this phantom had the same basic performances with European spine phantoms.This research is very important for the accurate measurement and value trace of the bone mineral densitometry of the standard phantom.

Keywords: DXA; Bone density; Standard phantom

2009, 35(1): 28-30,70  收稿日期: 2008-7-6;收到修改稿日期: 2008-9-22

基金项目: 科技部社会公益专项(2004DIAIJ003)

作者简介: 鄢铃(1969-),男,工程师,主要从事医学设备检测。

参考文献

[1] 刘忠厚,杨定焯,尚家芸.骨矿与临床[M].1版.北京:中国科学技术出版社,2006:303-308.
[2] Hermann K P,Geworsk L,Muth M,et al. Polyethylene-based water-equivalent phantom material for X-ray dosimetry at tube voltages from 10 to 100 kV[J]. Med.Phys.,1985,30(11):1195-1200.
[3] White D R,Martin R J,Darliaon R. Epoxy resin based tissue substitutes[J]. J.Radiol,1997(50):814-821.
[4] 杨定焯,尚家芸,宁文忠,等:DXA测量用校准脊椎体模的研制及临床测试[J].中国医学科学院学报,2003,25(3):262-266.
[5] Pearson J,Dequeker J,Henley M,et al.European semi-anthropomorphic spine phantom for calibration of bone densitometry:assessmet of precision,stability and accuracy[J].The European Quantitation of Osteoporosis Study Group.,Osteoporosis Int.,1995(5):174-184.
[6] 包安德,马海波,秦林林,等.不同双能X线密度仪的欧洲体模测定结果的比较研究[J].核电子学与探测技术,1999,19(5):351-354.
[7] The International Society for Clinical Densitopetry(ISCD).国际临床骨密度学会共识文件[S].2005版.程晓光译.中国骨质疏松杂志,2006,12(2):205-266.
[8] 杨勇,张圆月,刘志宏.医用诊断X射线剂量探测技术概述[J].中国测试技术,2005,31(4):94-96.
[9] 杨元第,孙志辉.医学计量的发展与展望[J].中国测试技术,2005,31(4):92-93.