单侧热绝缘热流计校准装置的研制

372    2024-03-22

¥0.50

全文售价

作者:王海涛, 文萌, 董亮, 孙云飞

作者单位:辽宁省计量科学研究院,辽宁 沈阳 110004


关键词:计量;热流计;校准装置;单侧热绝缘


摘要:

热流计是量化热能转移的仪器。为解决非对称平面热流计、热流传感器校准的难题,基于绝对法原理,对保护板法进行改进,建立一种单侧热绝缘的热流计校准装置,由主加热器、绝热层、保护加热器、功率输出系统、温度反馈调节系统和数据采集系统组成,装置可溯源至标准功率计。在400~1000 W/m2范围内对装置进行验证,测试结果与传统保护板法热流计校准装置标定结果基本一致。根据测试结果评估电流源偏差、背向和侧向热损失、蓄热损失、测量重复性、面积不一致、装载方式对测量结果的影响。经过实验和分析,该装置的不确定度可以达到3%以内。


Study and establishment of a single-side thermal insulation heat flow meter calibration device
WANG Haitao, WEN Meng, DONG Liang, SUN Yunfei
Liaoning Institute of Metrology, Shenyang 110004, China
Abstract: Heat flow meter is an instrument that quantifies the transfer of heat energy. In order to solve the problem of calibrating asymmetric planar heat flow meter and heat flow sensor, based on the principle of absolute method, the protective heat plate method was improved and a single-side thermally insulated heat flow meter calibration device was established, which was composed of main heater, insulation layer, protection heater, power output system, temperature feedback adjustment system and data acquisition system and the device can be traceable to a standard power meter. The device was validated in the range of 400~1000 W/m2, and the test results were in general agreement with the calibration results of the conventional protection plate method heat flow meter calibration device. The effects of current source deviation, backward and lateral heat loss, heat storage loss, measurement repeatability, area inconsistency and reinstallation on the measurement results were evaluated based on the test results. After the experiments and analysis, the uncertainty of the device can be achieved within 3%.
Keywords: metrology;heat flow meter;calibration device;single-side thermal insulation
2024, 50(3):131-136  收稿日期: 2021-10-20;收到修改稿日期: 2021-12-17
基金项目: 辽宁省检验检测认证中心青年人才创新创业项目(SC20190111)
作者简介: 王海涛(1985-),男,山东德州市人,高级工程师,硕士,主要研究方向为温度及热物性传感器的计量与研究。
参考文献
[1] 廖亚非, 张青文, 何容盛. 热流计的发展、国际领先技术与改进方向研究[J]. 重庆建筑大学学报, 2005(2): 84-87.
LIAO Y F, ZHANG Q W, HE R S.Research on the development, international leading technology and improvement direction of heat flow meters [J].Journal of Chongqing Jianzhu University, 2005 (2): 84-87.
[2] 胡芃, 陈则韶. 量热技术和热物性测定[M]. 合肥: 中国科学技术大学出版社. 2009: 31-32.
HU P, CHEN Z S. Calorimetry and thermophysical property determination [M]. Hefei: Press of University of Science and Technology of China, 2009: 31-32
[3] EVANGELISTI L, GUATTARI C, ASDRUBALI F. Comparison between heat-flow meter and Air-Surface Temperature Ratio techniques for assembled panels thermal characterization[J]. Energy and Buildings, 2019, 203: 109441.
[4] 李禹江, 童涛, 宫金林, 等. 发电机出线裸导体电磁-热防护研究[J]. 中国测试, 2019, 45(5): 60-65.
LI Y J, TONG T, GONG J L, et al. Research on electromagnetic thermal protection of bare conductors for generator outgoing lines [J]. China Measurement & Test, 2019, 45 (5): 60-65.
[5] 刘岩, 张伟. 基于热真空环境模拟试验技术的信号电缆热防护方案研究[J]. 电子测量技术, 2019, 42(9): 69-73.
LIU Y, ZHANG W. Research on thermal protection scheme for signal cables based on thermal vacuum environment simulation testing technology [J].Electronic Measurement Technology, 2019, 42 (9): 69-73.
[6] 王青, 江剑. 热流密度测试系统的设计与开发[J]. 国外电子测量技术, 2019, 38(6): 68-71.
WANG Q, JIANG J. Design and development of a heat flux density testing system [J].Foreign Electronic Measurement Technology, 2019, 38 (6): 68-71.
[7] PICCINI E, GUO S M, JONES T V. The development of a new direct-heat-flux gauge for heat-transfer facilities[J]. Measurement Science & Technology, 2000, 11(4): 342-349.
[8] 张俊锋, 苏健军, 姬建荣, 等. 热容式热流密度传感器的研制及应用[J]. 中国测试, 2018, 44(10): 125-129.
ZHANG J F, SU J J, JI J R, et al.Development and application of heat capacity heat flow density sensor [J].China Measurement & Test, 2018, 44 (10): 125-129.
[9] 于向阳, 姚凌虹, 赵时, 等. 基于红外特征的飞机导线绝缘层隐性故障在线诊断方法[J]. 电子测量与仪器学报, 2019, 33(10): 157-164.
YU X Y, YAO L H, ZHAO S, et al. An online diagnosis method for hidden faults in aircraft wire insulation layer based on infrared features [J].Journal of Electronic Measurement and Instrumentation, 2019, 33 (10): 157-164.
[10] FRANCESCO, PAOLO, D 'ALEO, et al. Design, calibration and testing of a thin film temperature gauge array for temperature and heat flux measurements in fluid mixing experiments[J]. Flow Measurement and Instrumentation, 2012, 24(1): 29-35.
[11] 周静雷, 张倩云, 李腾飞. 用于寿命试验仪的扬声器热传递参数测量方法[J]. 国外电子测量技术, 2019, 38(5): ;1-5.
ZHOU J L, ZHANG Q Y, LI T F. Measurement method for thermal transfer parameters of speakers used in life testing instruments [J].Foreign Electronic Measurement Technology, 2019, 38 (5): 1-5.
[12] 王玉刚, 轩诗瑶, 赵晓东, 等. 液冷散热器性能测试系统的研制[J]. 中国测试, 2020, 46(4): 97-101,142.
WANG Y G, XUAN S Y, ZHAO X D, et al.Development of a performance testing system for liquid cooled radiators [J].China Measurement & Test, 2020, 46 (4): 97-101,142.
[13] PONTE F D, MACCATO W. The calibration of heat flow meters[J]. Astm Special Technical Publication, 1980: 237-254.
[14] 秦永烈, 王耿, 赵学端. 热流计标定装置热损误差的计算与分析[J]. 上海机械学院学报, 1987(4): 33-38.
QIN Y L, WANG G, ZHAO X D. Calculation and analysis of heat loss error in heat flow meter calibration device [J].Journal of Shanghai Institute of Mechanical Engineering, 1987 (4): 33-38.
[15] 居滋培. 热流传感器特性和标定方法分析[J]. 传感技术学报, 1998(2): 32-36.
JU Z P. Analysis of characteristics and calibration methods of heat flow sensors [J].Journal of Transduction Technology, 1998 (2): 32-36.
[16] ASTM C177 Standard Test Method for Steady-State Heat Flux Measurements and Thermal Transmission Properties by Means of the Guarded-Hot-Plate Apparatus [S]. ASTM, 2019.
[17] 杨家健, 原遵东, 韩蕊, 等. 热流计校准装置测量控制系统的建立[J]. 电子测量与仪器学报, 2011, 25(7): 620-625.
YANG J J, YUAN Z D, HAN R, et al. The establishment of a measurement control system for the calibration device of a heat flow meter [J].Journal of Electronic Measurement and Instrument, 2011, 25 (7): 620-625.